Case 2
- DOI:
- 10.1093/med/9780198727774.003.0002
Introduction to the case
An 82-year-old woman had been implanted with a single-chamber ventricular pacemaker 6 years ago for atrial fibrillation (AF) with symptomatic bradycardia. She presented to the hospital with shortness of breath and twitching of the left pectoral region. Her pacemaker follow-up, performed 6 months ago, had been completely normal. The results of the device follow-up performed at that time are shown in Table 2.1, and the electrocardiogram (ECG) recorded at admission is shown in Figure 2.1.
Table 2.1 Device settings and tested parameters
Programming |
|
---|---|
Mode |
VVIR |
Rate |
60-110bpm |
Pacing output (bipolar) |
2.5V/0.4ms |
Sensitivity setting (bipolar) |
2.8mV |
Tests |
|
Battery voltage |
2.71V |
Battery impedance |
1200 ohms |
Lead impedance (bipolar) |
625 ohms |
Capture threshold (bipolar) |
1.2V/0.4ms |
Sensing threshold (bipolar) |
6.8mV |
The pacing spikes in Figure 2.2 occur at exactly 60bpm (the baseline rate), with all spikes showing ventricular undersensing and non-capture. This should not be confused with non-pacing, which refers to the lack of delivery of a pacing spike where one is expected.
-
➊ This beat could be interpreted as normal device function with pseudofusion, as the spike occurs almost simultaneously with the QRS complex. With pseudofusion, the pacing spike occurs at, or shortly after, QRS onset, but before the depolarization wavefront has reached the lead dipole. The pacing spike occurs during the myocardial refractory period and therefore does not capture the ventricle.
-
➋ The ventricular spike occurs on the T-wave which may be potentially pro-arrhythmic, but this is not the case here, as there is no capture.
Subclavian crush
A chest X-ray showed complete lead section due to subclavian crush (Figure 2.3A). A chest X-ray had been performed shortly before the previous device follow-up by the patient’s general practitioner for cough and already showed signs of lead damage (Figure 2.3B).
This case illustrates how normal parameters (including impedance) do not rule out damage to a lead. This is of particular relevance in pacemaker-dependent patients who may be experiencing symptoms compatible with device dysfunction and in whom a chest X-ray and Holter recording should be performed if device interrogation does not elucidate the problem.
Subclavian crush may result from medial subclavian puncture and may be avoided by favouring axillary vein puncture or cephalic vein cutdown for lead implantation.